Autoclave Steriale 1500 Liters

Class B autoclave for up to 11x 60 liter Steriale containers, with integrated GPRS system.

Presentation

STERIALE-1500 is an automatic class B autoclave with integrated GPRS system that operates at high temperature and pressure, offering a fast sterilization process through the use of steam. It is used with Steriale’s 30 and 60 liter containers, and all sharps containers.

Up to 11x 60-liter containers can be sterilized at a time.

Intended Use

This autoclave has been specifically designed for the sterilization of infectious biological waste in healthcare facilities. It is ideal for hospitals, clinics, laboratories and other medical centers that need to manage their waste safely and efficiently.

It must be used in conjunction with the special Steriale System supplies and containers.

Certifications

  • ISO 9001
  • ISO 13485
  • ISO 14001
  • CE
  • PED – Pressure Vessel

Features

  • Class B autoclave with vacuum pump and configurable pre-vacuum cycles.
  • Internal chamber allows sterilization of one or more containers at the same time.
  • GPRS module for data transmission and traceability via QR code reader.
  • High precision scales on the feet for waste weighing.
  • Presetting includes 3 sterilization cycles: Solid Cycle, Liquid Cycle and Leak Test.

Steriale Application

  • The GPRS module transmits all sterilization process data to the Steriale Software.
  • The control system automatically records the date and time at which sterilization starts and ends.
  • Includes SIM Card and SMA Antenna for internet connection.

Traceability

  • High performance barcode scanner module with fast reading. It is capable of reading codes in low light or low resolution conditions.
  • It can read ordinary barcodes, 2D barcodes and barcodes on mobile device displays.

HEPA filter

Air extraction system equipped with HEPA antibacterial filter.

Automatic Residue Weighing

High precision scale on the feet records the weighing of the waste before starting the cycle and also at the end of the cycle, allowing to identify how much weight has been lost during the cycle. All the information is sent to the Steriale Software.

Installation of Water Tanks

The autoclaves need to be installed with 2 external water tanks.

  • Tank 1 is the temporary waste tank, where the water discarded by the autoclave is first directed into the tank (1A), passed through a coil (heat sink) and then discharged directly to the appropriate municipal drain (1B).
  • Tank 2 is the water recirculation tank for the vacuum pump. It is a water that must be changed every X days, defined during the installation of the system and notified to the health facility. This water is clean, obtained from reverse osmosis and enters through inlet 2A, feeds the vacuum pump (2C) and recirculates, returning to the tank again through outlet 2B.

Technical Information

STERIALE-1500
Volume1500 Liters
Cargo Capacity11 Containers
TypeClass B
Camera Size (W × H × D) mm680 × 1180 × 1870
Overall Size (L × W × H) mm1450 × 1990 × 2080
Net Weight~ 2,130 kg
Power SupplyAC 380V / 50 Hz
Nominal Power (KW)100
Temperature Range (adjustable per cycle)115 ~138 ºC
Sterilization Time (adjustable)0~99 minutes
Chamber Material304 Stainless Steel
Design Pressure-0.1 / 0.3 MPa
Design Temperature144ºC
Vacuum Pump Technical InformationMaximum Vacuum: ≤ -90 kPa
Non-linear error margin of weight sensors:±0.05% R.O.
Display8.4 inch LCD screen
Preconfigured LanguagesEnglish, Spanish, Mandarin and Portuguese
OperationTouch Screen
Door opening and closing methodAutomatic Door
Pressure gauge2
Access ManagementYou can manage operators with multilevel authority by setting passwords.
Safety SystemsAutomatic overheating protection device, bidirectional safety locking device for door pressure and temperature, safety valve for automatic overpressure relief, short-circuit protection and overvoltage protection.
Digital Communication MethodGPRS Steriale System with SIM card or Wi-Fi
QR Code ScannerIntegrated RS232 communication scanner
Weight Sensors4 units with full bridge connection
Antenna Connector TypeSMA
Steam Generator
Water Quality Detection System
Preconfigured sterilization cyclesSolid Residues, Liquid Residues, Leak Test
Can be installed on double bedplate

No

Sterilization Cycles

Infectious Waste CycleLeak Test CycleLiquid Waste Cycle
Intended useSterilization of mostly dry infectious biological waste.Daily leak test to ensure that the autoclave does not leak.Sterilization of mostly liquid infectious biological waste.
Total Time50 minutes15 minutes80-90 minutes
Sterilization Time8 minutes4 minutes10 minutes
Drying Time5 minutes~5 minutes
Temperature134ºC134ºC134ºC
Pressure2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm
All cycles come from the factory and can be configured to best suit the local legislation of each country or region.