Presentation
STERIALE-80 is an automatic class B autoclave with integrated GPRS system that operates at high temperature and pressure, offering a fast sterilization process through the use of steam. It is used with Steriale’s 30 and 60 liter containers, and all sharps containers.
Intended Use
This autoclave has been specifically designed for the sterilization of infectious biological waste in healthcare facilities. It is ideal for hospitals, clinics, laboratories and other medical centers that need to manage their waste safely and efficiently.
It must be used in conjunction with the special Steriale System supplies and containers.
Certifications
- ISO 9001
- ISO 13485
- ISO 14001
- CE
- PED – Pressure Vessel
Features
- Class B autoclave with vacuum pump and configurable pre-vacuum cycles.
- Internal chamber allows sterilization of one or more containers at the same time.
- GPRS module for data transmission and traceability via QR code reader.
- High precision scales on the feet for waste weighing.
- Presetting includes 3 sterilization cycles: Solid Cycle, Liquid Cycle and Leak Test.


Steriale Application
- The GPRS module transmits all sterilization process data to the Steriale Software.
- The control system automatically records the date and time at which sterilization starts and ends.
- Includes SIM Card and SMA Antenna for internet connection.

Traceability
- High performance barcode scanner module with fast reading. It is capable of reading codes in low light or low resolution conditions.
- It can read ordinary barcodes, 2D barcodes and barcodes on mobile device screens.

HEPA filter
Air extraction system equipped with HEPA antibacterial filter.

Automatic Residue Weighing
High precision scale on the feet records the weighing of the waste before starting the cycle and also at the end of the cycle, allowing to identify how much weight has been lost during the cycle. All the information is sent to the Steriale Software.
Installation of Water Tanks
The autoclaves need to be installed with 2 external water tanks.
- Tank 1 is the temporary waste tank, where the water discarded by the autoclave is first directed into the tank (1A), passed through a coil (heat sink) and then discharged directly to the appropriate municipal drain (1B).
- Tank 2 is the water recirculation tank for the vacuum pump. It is a water that must be changed every X days, defined during the installation of the system and notified to the health facility. This water is clean, obtained from reverse osmosis and enters through inlet 2A, feeds the vacuum pump (2C) and recirculates, returning to the tank again through outlet 2B.

Technical Information
STERIALE-80 | |
---|---|
Volume | 80 Liters |
Load Capacity | 1 Container |
Type | Class B |
Chamber Size (W × H × D) mm | Ø385 × 717 |
Overall Size (L × W × H) mm | 1040 × 580 × 580 |
Net Weight | ~ 185 kg |
Power Supply | AC 220V / 50 Hz |
Nominal Power (KW) | 2.6 |
Temperature Range (adjustable per cycle) | 105 ~138 ºC |
Sterilization Time (adjustable) | 0~99 minutes |
Chamber Material | 304 Stainless Steel |
Design Pressure | -0.1 / 0.3 MPa |
Design Temperature | 144ºC |
Vacuum Pump Technical Information | Maximum Vacuum: ≤ -90 kPa |
Non-linear error margin of weight sensors: | ±0.05% R.O. |
Display | 3.2 inch LCD screen |
Preconfigured Languages | English, Spanish, Mandarin and Portuguese |
Operation | 5 sensor keys |
Door opening and closing method | Manual + Automatic door |
Pressure gauge | 2 |
Access Management | You can manage operators with multilevel authority by setting passwords. |
Safety Systems | Automatic overheating protection device, bidirectional safety locking device for door pressure and temperature, safety valve for automatic overpressure relief, short-circuit protection and overvoltage protection. |
Digital Communication Method | GPRS Steriale System with SIM card or Wi-Fi |
QR Code Scanner | Integrated RS232 communication scanner |
Weight Sensors | 4 units with full bridge connection |
Antenna Connector Type | SMA |
Steam Generator | ✔ |
Water Quality Detection System | ✔ |
Preconfigured sterilization cycles | Solid Residues, Liquid Residues, Leak Test |
Can be installed on double bedplate | ✔ |
Sterilization Cycles
Infectious Waste Cycle | Leak Test Cycle | Liquid Waste Cycle | |
---|---|---|---|
Intended use | Sterilization of mostly dry infectious biological waste. | Daily leak test to ensure that the autoclave does not leak. | Sterilization of mostly liquid infectious biological waste. |
Total Time | 50 minutes | 15 minutes | 80-90 minutes |
Sterilization Time | 8 minutes | 4 minutes | 10 minutes |
Drying Time | 5 minutes | ~ | 5 minutes |
Temperature | 134ºC | 134ºC | 134ºC |
Pressure | 2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm | 2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm | 2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm² 2.4 kgf/cm |
All cycles come from the factory and can be configured to best suit the local legislation of each country or region. |