Presentation
STERIALE-45 is an automatic autoclave that operates at high temperature and pressure, offering a rapid sterilization process using steam. It is designed for the sterilization of infectious biological waste. It must be used in conjunction with the special Steriale System supplies and containers.
Certifications
- MDD 93/42/EEC amended by 2007/47/EC PED 97/23/EC
- IEC 61010-1:2010
- IEC 61010-2-040:2015
- EN 13060:2014
- EN 61326:2005
Features
- Class B autoclave with vacuum pump and configurable pre-vacuum cycles.
- Internal chamber allows sterilization of one or more containers at the same time.
- GPRS module for data transmission and traceability via QR code reader.
- High precision scales on the feet for waste weighing.
- No need to press any button; when the door is closed, the cycle starts automatically


Steriale Application
- The GPRS module transmits all sterilization process data to the Steriale Software.
- The control system automatically records the date and time at which sterilization starts and ends.
- Includes SIM Card and SMA Antenna for internet connection.

Traceability
- High performance barcode scanner module with fast reading. It is capable of reading codes in low light or low resolution conditions.
- It can read ordinary barcodes, 2D barcodes and barcodes on mobile device screens.

HEPA filter
Air extraction system equipped with HEPA antibacterial filter.

Automatic Residue Weighing
High precision scale on the feet records the weighing of the waste before starting the cycle and also at the end of the cycle, allowing to identify how much weight has been lost during the cycle. All the information is sent to the Steriale Software.
Installation of Water Tanks
The autoclaves need to be installed with 2 external water tanks.
- Tank 1 is the temporary waste tank, where the water discarded by the autoclave is first directed into the tank (1A), passed through a coil (heat sink) and then discharged directly to the appropriate municipal drain (1B).

Technical Information
STERIALE-45 | |
---|---|
Volume | 45 Liters |
Type | Class B |
Chamber Size (W × H × L) mm | Ø319 × 550 |
Overall Size (L × W × H) mm | 865 × 506 × 536 |
Net Weight | ~ 105 kg |
Power Supply | AC 230V ~ 50/60 Hz |
Input Power (VA) | 3300VA |
Temperature Range | 121 ~134 ºC |
Chamber Material | 304 Stainless Steel |
Design Pressure | -0.1 / 0.3 MPa |
Design Temperature | 140ºC |
Accuracy Temperature | 0.1ºC |
Accuracy Pressure | 1 Kpa |
Non-linear error margin of weight sensors: | ±0.05% R.O. |
Display | LED Display |
Operation | 6 sensor keys |
Door opening and closing method | Automatic Door |
Digital Communication Method | GPRS Steriale system with SIM card or Wi-Fi. |
QR Code Scanner | Integrated RS232 communication scanner |
Weight Sensors | 4 units with full bridge connection |
Antenna Connector Type | SMA |
Steam Generator | Yes ✔ |
Water Quality Detection System | Yes ✔ |
Preconfigured sterilization cycles | Infectious Residue and Leak Test |
Cargo Capacity | 1 container of 30 liters |
Can be installed on double bedplate | Yes ✔![]() Example of a double bed installation |
Sterilization Cycles
Infectious Waste Cycle | Leak Test Cycle | |
---|---|---|
Intended use | Sterilization of mostly dry infectious biological waste. | Daily leak test to ensure that the autoclave does not leak. |
Total Time | 50 minutes | 15 minutes |
Sterilization Time | 8 minutes | 4 minutes |
Drying Time | 2 minutes | ~ |
Temperature | 134ºC | 134ºC |
Pressure | 2.4 bar | 2.4 bar |
All cycles come from the factory and can be configured to best suit the local legislation of each country or region. |